当前位置: 首页 > 文章 > 电感耦合等离子体发射光谱法(ICP-AES)测定污水中总磷 上海农业学报 2005,21 (2) 16-18
Position: Home > Articles > Determination of Total Phosphorus in Sewage by ICP-AES Acta Agriculturae Shanghai 2005,21 (2) 16-18

电感耦合等离子体发射光谱法(ICP-AES)测定污水中总磷

作  者:
张栩;周德兴;袁大伟;孙锡娟;何七勇;雷萍
单  位:
上海市农业科学院环境科学研究所
关键词:
ICP-AES法;污水;磷
摘  要:
采用美国热电公司TraceScan型端视ICP AES光谱仪,对测定污水中总磷的最佳测试条件的选择进行了探讨,并用所选取的最佳测试条件进行测试。结果表明,本方法简便,可以直接上机进行测定;快速而准确,相对误差为0%~2.00%;精密度好,相对标准偏差为0.3%~4.2%(n=3);回收率在104.00%~118.00%之间,检出限0.015mg/L。
译  名:
Determination of Total Phosphorus in Sewage by ICP-AES
作  者:
ZHANG Xu, ZHOU De-xing, YUAN Da-wei, SUN Xi-juan, HE Qi-yong, Lei Ping(Environmental Sciences Research Institute, Shanghai Academy ofAgricultural Sciences, Shanghai 201106,China)
关键词:
ICP-AES; Sewage; Phosphorus
摘  要:
The optimal condition of testing total phosphorus amount of sewage was studied and the tests were carried out according to the selected optimal condition by means of Trace Scan type ICP-AES spectrograph made by ThermoElemental Company in the USA. The test results showed that this method of testing total phosphorus amount of sewage by means of ICP-AES technology is not only simple and convenient but also fast and accurate. Its relative error is between 0%~2.00%, and relative standard deviation is between 0.3%~4.2%(n=3). Its recovery rate is between 104.00%~118.00%. Its detection limit is 0.015mg/L.