当前位置: 首页 > 文章 > 高效液相色谱法同柱分析草甘膦和双甘膦 农药 2009,48 (6) 47-48
Position: Home > Articles > Analysis of Glyphosate and N-(Phosphonomethyl)iminodiacetic Acid at the Same Column by HPLC Agrochemicals 2009,48 (6) 47-48

高效液相色谱法同柱分析草甘膦和双甘膦

作  者:
张丽娟;韦少平;李致宝;许朝芳;翁德洪;柯敏
单  位:
广西化工研究院
关键词:
草甘膦;双甘膦;高效液相色谱;定量
摘  要:
采用高效液相色谱法,以磷酸二氢钾缓冲溶液作为流动相,Hypersil SAX色谱柱,流速0.8mL/min,紫外检测器波长195nm下同柱分析草甘膦和双甘膦的含量。结果表明草甘膦和双甘膦的线性相关系数为0.9999和0.9984,标准偏差为0.0899和0.7701,变异系数为0.1620%和0.9660%。
译  名:
Analysis of Glyphosate and N-(Phosphonomethyl)iminodiacetic Acid at the Same Column by HPLC
作  者:
ZHANG Li-juan, WEI Shao-ping, LI Zhi-bao, XU Chao-fang, WENG De-hong, KE Min (Guangxi Research Institute of Chemical Industry, Nanning 530001, China)
关键词:
glyphosate; N-(phosphonomethyl)iminodiacetic acid; HPLC; quantitative
摘  要:
The quantitative analysis of glyphosate and N-(phosphonomethyl)iminodiacetic acid was conducted using HPLC with potassium dihydrogen phosphate buffer solution as mobile phase, Hypersil SAX column, flow rate: 0.8 mL/min, using UV detector at 195 nm wavelength to quantitative analysis glyphosate and N-(phosphonomethyl)iminodiacetic acid at the same column. The results showed that linear correlation for glyphosate and N-(phosphonomethyl)iminodiacetic acid were 0.999 9 and 0.998 4, the standard deviation were 0.089 9 and 0.770 1, the variation coefficient were 0.162 0% and 0.966 0%, respectively.